ANALISIS HASIL Quality Control PEMERIKSAAN DARAH RUTIN MENGGUNAKAN Hematology Analyzer Sysmex XN-350 DI RSUD SITI FATIMAH PROVINSI SUMATERA SELATAN

Mitha Novitha, P07134321008 (2022) ANALISIS HASIL Quality Control PEMERIKSAAN DARAH RUTIN MENGGUNAKAN Hematology Analyzer Sysmex XN-350 DI RSUD SITI FATIMAH PROVINSI SUMATERA SELATAN. Diploma thesis, Poltekkes Kemenkes Yogyakarta.

[img] Text
Awal.pdf

Download (822kB)
[img] Text
Abstract.pdf

Download (214kB)
[img] Text
Chapter 1.pdf

Download (328kB)
[img] Text
Chapter 2.pdf

Download (575kB)
[img] Text
Chapter 3.pdf

Download (450kB)
[img] Text
Chapter 4.pdf
Restricted to Registered users only

Download (360kB)
[img] Text
Conclusion.pdf
Restricted to Registered users only

Download (387kB)
[img] Text
References.pdf

Download (405kB)
[img] Text
Appendices.pdf
Restricted to Registered users only

Download (2MB)

Abstract

Quality Control merupakan proses yang dilakukan untuk mengevaluasi pengujian dan memastikan bahwa sistem mutu berjalan baik dengan tujuan menjamin hasil pemeriksaan laboratorium, meminimalkan penyimpangan serta mengetahui sumber dari kesalahan sehingga diperoleh hasil pemeriksaan yang tepat dan dapat dipercaya. Mengetahui hasil quality control pemeriksaan darah rutin menggunakan hematology analyzer XN-350 di RSUD Siti Fatimah Provinsi Sumatera Selatan. Penelitian observasional analitik, cross sectional. Penelitian dilaksanakan pada bulan Desember 2022. Populasi penelitian adalah data QC hematology analyzer XN-350, berjumlah 450 data dalam satu bulan, dianalisis menggunakan rumus stastistical quality control. QC low level parameter Hemoglobin CV 1,4%, (d%) 1,6%, sigma 4,9, Eritrosit CV 1,29%, (d%) 0,11%, sigma 4,57, Leukosit CV 3,14%, (d%) 0,36%, sigma 4,66, Hematokrit CV 1,1%, (d%) 0,30%, sigma 5,1, Trombosit CV 2,5%, (d%) 10,4%, sigma 5,8. QC normal level parameter Hemoglobin CV 0,8%, (d%) 1,5%, sigma 7,2, Eritrosit CV 0,61%, (d%) 0,79%, sigma 8,56, Leukosit CV 1,97%, (d%) 0,48%, sigma 7,38, Hematokrit CV 0,6%, (d%) 0,3%, sigma 10,1, Trombosit CV 2,2%, (d%) 7,2%, sigma 8,0. QC high level parameter Hemoglobin CV 0,7%, (d%) 2,7%, sigma 5,9, Eritrosit CV 0,75%, (d%) 1,17%, sigma 6,44, Leukosit CV 1,20%, (d%) 0,30%, sigma 12,24, Hematokrit CV 0,6%, (d%) 1,3%, sigma 7,4, Trombosit CV 1,4%, (d%) 7,2 %, sigma 12,9. Hasil QC low level, normal level dan high level parameter eritrosit, leukosit, hematokrit mendapatkan presisi, akurasi tinggi dan six sigma level good performance-world class performance. Sedangkan parameter hemoglobin dan trombosit mendapatkan presisi tinggi, akurasi rendah dan six sigma level good performance-world class performance.

Item Type: Thesis (Diploma)
Subjects: R Medicine > RZ Other systems of medicine